Apolipoprotein A-1 (ApoA1)
Used for the determination of serum Apolipoprotein A-1 on the Beckman Coulter UniCel DxC600 Synchron Clinical System.
Instrument: Beckman Coulter UniCel DxC600
Specimen Type: Serum
Volume per test: 15uL (min volume 125uL)
Principle:
ApoA reagent is used to measure ApoA1 concentration by a turbidimetric method. The Synchron DxC System automatically proportions the sample and reagent volumes into the cuvette at a ratio of 1:38. ApoA1 combines with specific antibody to form an insoluble antigen-antibody complex, resulting in turbidity. The change in absorbance at 340 nanometers is directly proportional to the concentration of ApoA1 in the sample and used by the system to calculate and report ApoA1 concentration.
Testing Range: 0.25-3.00 g/L